5-5

Troubleshooting

5

Electrical

Electrical

Input functions not operating

• Make sure connections into terminal block

connector are secure.

• Make sure terminal block connector is secure in

input/output port.

• Check connections for proper pin assignments.

• Be sure pins from external devices are assigned

correctly.

• Check polarity of input. Inputs should be a

contact closure. If not, it must be TTL level

(logic 0 activates).

• Confirm that source supplying input to liquid

handler is working.

Output functions not operating

• Make sure connections into terminal block

connector are secure.

• Make sure terminal block connector is secure in

the input/output port.

• Check connections for proper pin assignments.

• Output from liquid handler should be compatible

with device to which it is interfaced. Outputs

are contact closures.