Installing different type of probe

To install a replacement probe of a different type

than is currently installed, you may want to obtain

a new probe guide insert for precise XY probe

accuracy.

1 Remove the transfer tubing’s 1/4"-28 fitting

connected to the top of the isolation probe

holder.

2 Grasp the current probe and push it up through

the top of the isolation probe holder.

3 Remove the current probe guide insert by

turning it counterclockwise. Then place the

new probe guide insert into probe guide and

secure it by turning it clockwise.

4 Insert the new probe into the top of the

isolation probe holder and pull it through the

holder and the new probe guide insert until the

tip of the probe is in the probe guide.

5 Replace and tighten the 1/4"-28 fitting.