Liquid detector, Liquid detector -9 – Gilson 215 User Manual

Page 49

Advertising
background image

3-9

Operation

3

Liquid Detector

The Liquid Detector tab allows you to adjust the

default sensitivity for liquid detection.

Note: Gilson control software (such as 735 Sampler

Software) will override these settings.

Raising the percentage suppresses false liquid

detection while lowering the percentage increases

sensitivity for situations where liquid is harder to

detect. By lowering the probe so it touches the

liquid for one of your samples, you can manually

test the current sensitivity of the liquid detection. In

this manner, you can observe the amount of

change you might expect with each sample.

The factory default setting is 10% for each probe.

Configuring the Liquid Handler