GSIOC Commands

Ksxxx

B

Sets liquid level sensing sensitivity where:

s - selected probe (a, b, c, or d)

xxx - Value from 0 to 255; 0 is the most sensitive, 255 is

the least sensitive.

If s is not specified, the default is abcd.

K

I

Reads liquid level sensing sensitivity. Returns

“aaa,bbb,ccc,ddd” where:

aaa - sensitivity of probe A

bbb - sensitivity of probe B

ccc - sensitivity of probe C

ddd - sensitivity of probe D

Possible sensitivity range is from 0 to 255 for each probe; 0

is the most sensitive, 255 is the least sensitive.

M

I

Reads X, Y, Z motor status and dilutor status. Returns

“xyzp”. For each status, you’ll see U for unpowered, P for

powered, R for running, or E for error.

Note: Because the Quad-Z uses an external dilutor, the

dilutor status is always unpowered (U).

m

I

Read X, Y, Z motor status and width of the probes. Returns

“xyabcdw” where:

x = X motor

y = Y motor

a = Z motor on probe A

b = Z motor on probe B

c = Z motor on probe C

d = Z motor on probe D

w = pitch motor

For each status, you’ll see U for unpowered, P for powered,

R for running, or E for error.

M

B

Start to move probes to the Z-height positions (set using the

buffered T command, see page

C-8

).

m

B

Start to move probes to the Z-height positions (set using

the buffered T command, see page

C-8

) with liquid level

sensing on.