Gilson 215 User Manual

Page 38

Advertising
background image

2-22

Installation

2

Baud rate/mode

As a default, the baud rate/mode is set to 6,

indicating that the liquid handler is set for a baud

rate of 19200 and is a master device.

You’ll need to change the selection to 0 (zero) if the

liquid handler is connected via the GSIOC to a

Gilson system and is being controlled by the 506C

System Interface from Gilson control software. A

setting of 0 indicates the liquid handler is a slave

device and the baud rate is being clocked

externally.

Note that other selections are available if the liquid

handler is being controlled by non-Gilson

applications. Refer to the table shown on the liquid

handler’s rear panel; see page

2-16

.

To change the baud rate/mode:

1 Gently insert a small flat blade screwdriver into

the SW2 selector on the rear panel and turn it.

2 Align the white dot with one of the indicated

numbers.

Electrical Connections