Specifications – Hach-Lange SI794 C User Manual User Manual

Page 100

Advertising
background image

HOLD input

Galv. separated (OPTO coupler)

Function

Switches transmitter to HOLD mode

Switching voltage

0 ... 2 V (AC/DC)

inactive

10 ... 30 V (AC/DC) active

CONTROL input

Galv. separated (OPTO coupler)

Function

Switch-over to second parameter set

Switching voltage

0 ... 2 V (AC/DC)

Parameter set 1

10 ... 30 V (AC/DC) Parameter set 2

Output 1

0/4 to 20 mA, max. 10 V, floating

(galv. connected to output 2)

Process variable

*

Conductivity, resistivity, concentration,

or salinity

Characteristic

Linear or logarithmic

Overrange

*

22 mA in the case of error messages

Output filter

*

Low-pass, filter time constant 0 ... 120 s

Meas. error

1

)

< 0.3 % current value + 0.05 mA

Start/end of scale

As desired within range

Minimum span

LIN:

5 % of selected range

LOG:

1 decade

Output 2

0/4 to 20 mA, max. 10 V, floating

(galv. connected to output 1)

Process variable

Temperature

Overrange

*

22 mA in the case of temp error messages

Output filter

*

Low-pass, filter time constant 0 ... 120 s

Meas. error

1

)

< 0.3 % current value + 0.05 mA

Start/end of scale

*

-20 ... 200 °C / -4 ... 392 °F

Span allowed

20 ... 320 K (36 ... 608 °F)

Alarm contact

Relay contact, floating

Contact ratings

AC< 250 V / < 3 A / < 750 VA

DC< 30 V / < 3 A / < 90 W

Contact response

N/C (fail-safe type)

Alarm delay

0000 ... 0600 s

100

Specifications