Configuration

Alarm settings

Select Sensocheck

Delay

LED in HOLD mode

Code Display

Menu group

Select menu item

Alarm settings

AL .

Exit:

Press conf, then enter.

15

16

Alarm

Alarm contact

The alarm contact is closed during normal operation (N/C).

It opens in the case of alarm or power outage.

As a result, a failure message is provided even in the case

of line breakage (fail-safe behavior). For contact ratings,

see Specifications.

Error messages can also be signaled by a 22 mA output

current (see Pg 45,

The operating behavior of the alarm contact is shown on

Pg 90 .

The alarm delay acts on the LED, the 22 mA signal, and

the alarm contact.

51, 88).