Calibration with calibration solution

Input of temperature-corrected value of calibration solution with

simultaneous display of cell constant

Press cal, enter code: 0110

Select using 4, edit number

using5, proceed with enter.

Immerse sensor in calibration

solution.

Determine the temperature-

corrected conductivity value of

the calibration solution from

the corresponding table

(see Pg 104).

The transmitter is in

the Hold mode.

If an invalid code is

entered, the trans-

mitter returns to

measuring mode.

Ready for calibration

Remove and clean sensor

Display (3 sec)

Enter value of calibration

solution.

Select using 4 key,

edit number using 5 key.

Press enter to confirm the

calibration data.

The cell constant

and temperature

are alternately

displayed in the

lower display

during the input.