Display
Action
Remark
When there has
not been an entry
for 6 sec, the lower
display alternately
shows the cell
constant and
temperature value.
Calibration with calibration solution
Input of temperature-corrected value of calibration solution with
simultaneous display of cell constant
Press cal, enter code: 0110
Select using 4, edit number
using5, proceed with enter.
Immerse sensor in calibration
solution.
Determine the temperature-
corrected conductivity value of
the calibration solution from
the corresponding table
(see Pg 104).
The transmitter is in
the Hold mode.
If an invalid code is
entered, the trans-
mitter returns to
measuring mode.
Ready for calibration
Remove and clean sensor
Display (3 sec)
Enter value of calibration
solution.
Select using 4 key,
edit number using 5 key.
Press enter to confirm the
calibration data.
The cell constant
and temperature
are alternately
displayed in the
lower display
during the input.