Temperature in °C Conductivity in µS/cm Temperature in °C Conductivity in µS/cm
0
0.6
55
2.1
5
0.8
60
2.2
10
0.9
65
2.4
15
1.0
70
2.5
20
1.1
75
2.7
25
1.3
80
2.7
30
1.4
85
2.7
35
1.5
90
2.7
40
1.7
95
2.9
45
1.8
100
3.1
50
1.9
Temperature/conductivity table as per USP
USP factor
Relay 1 active
Cond>USP (Table)
10 %
100 %
Reduced
USP limit value
USP limit value
see table above
Err 01
Conductivity within limits of USP function
Limit values for USP function
Limit contact response U1 (Relay 1) and U2 (Relay 2)
Relay 2 active