Default settings of parameter sets – Hach-Lange SI794 C User Manual User Manual

Page 69

Advertising
background image

Two complete parameter sets are stored in the EEPROM . As

delivered, the two sets are identical but can be edited differently.

Note:

Fill in your configuration data on the following pages.

Code. Parameter

Default setting

o1. Sensor selection * 2-EL

o1. Process variable

000.0 mS

o1. Conc solution

-01-

o1. 0/4-20 mA

4-20 mA

o1. Characteristic

LIN

o1. Current start (LIN) 000.0 mS

o1. Current end (LIN)

100.0 mS

o1. Current start (LOG) 0.1 mS

o1. Current end (LOG) 100 mS

o1. Filter time

0 s

o1. 22mA signal

OFF

o1. HOLD response

Last

o1. Fix current

021.0 mA

o2. Unit °C / °F

°C

o2. Temp probe *

Pt100

o2. 0/4 ...20mA

4-20 mA

o2. Current start

000.0 °C

o2. Current end

100.0 °C

o2. Filter time

0 s

o2. 22mA signal

OFF

o2. HOLD response

Last

o2. Fix current

021.0 mA

tc. Temp compensation OFF

tc. Temp coefficient

02.00%/K

AL. Sensocheck

OFF

AL. Alarm delay

0010 sec

AL. LED Hold

OFF

Code. Parameter

Default setting

L1. Contact function

Lo

L1. Contact response

N/O

L1. Setpoint

000.0 mS

L1. Hysteresis

001.0 mS

L1. Delay

0010 sec

L2. Contact function

Hi

L2. Contact response

N/O

L2. Setpoint

100.0 mS

L2. Hysteresis

001.0 mS

L2. Delay

0010 sec

Ct. Setpoint

050.0 mS

Ct. Neutral zone

001.0 mS

Ct. P action

0100 %

Ct. I action

0000 sec

Ct. D action

0000 sec

Ct. PLC/PFC controller PLC

Ct. Pulse length

0010 sec

Ct. Pulse frequency

0060 /min

Ct. Hold response

Last

U1 USP factor

100 %

U1 Contact response N/O

U1 Delay

0000 sec

U2 Contact response N/O

U2 Delay

0000 sec

Cn. Rinse/ PSet2 *

rinse

Cn. Rinsing interval *

000.0 h

Cn. Rinse duration *

0060 sec

Cn. Contact type *

N/O

Default settings of parameter sets

69

*) These parameters are only edited in parameter set 1.

They are valid for both parameter sets.