Specifications – Hach-Lange SI794 C User Manual User Manual

Page 98

Advertising
background image

98

Conductivity input

Input for 2-electrode/4-electrode sensors

Effective ranges

Conductivity

0.2 µS · c ... 1000 mS · c

Ranges

*

Conductivity

0.000 ... 9.999 µS/cm

000.0 ... 99.99 µS/cm

000.0 ... 999.9 µS/cm

0000 ... 9999 µS/cm

0.000 ... 9.999 mS/cm

00.00 ... 99.99 mS/cm

000.0 ... 999.9 mS/cm

0.000 ... 9.999 S/m

00.00 ... 99.99 S/m

Resistivity

00.00 ... 99.99 MΩ · cm

Concentration

0.00 ... 9.99% by wt

Salinity

0.0 ... 45 ‰ (0 ... 35 °C)

Response time (T

90

)

< 1 s (Sensocheck off)

< 3 s (Sensocheck on)

Measurement error

1,2,3)

< 1 % meas. val. +0.4 µS · c

Concentration determination
Operating modes

*

-01- NaCl 0.00 ... 9.99% by wt (0 ... 60 °C)

-02- HCl

0.00 ... 9.99% by wt (-20 ... 50 °C)

-03- NaOH 0.00 ... 9.99% by wt (0 ... 100 °C)

-04- H

2

SO

4

0.00 ... 9.99% by wt (-17 ... 110 °C)

-05- HNO

3

0.00 ... 9.99% by wt (-17 ... 50 °C)

See graphs in the Appendix Pg 106 et seq.

Sensor standardization
Operating modes

• Input of cell constant with simultaneous

display of conductivity and temperature

• Input of conductivity of calibration solution

with simultaneous display of cell constant and

temperature

• Product calibration

• Temperature probe adjustment

Permssible cell constant

00.0050 ... 19.9999 cm

-1

Specifications