Code display action choices o1, Concentration measurement – Hach-Lange SI794 C User Manual User Manual

Page 37

Advertising
background image

37

Code Display

Action

Choices

o1 .

Only with 00.00 % conc:

select the process solution

Select using 4 key.

-01- NaCl (0.00 ... 9.99 % by wt)

(0 ... 120 °C)

-02- HCl (0.00 ... 9.99 % by wt)

(-20 ... 50 °C)

-03- NaOH (0.00 ... 9.99 % by wt)

(0 ... 100 °C)
-04- H

2

SO

4

(0.00 ... 9.99 % by wt)

(-17 ... 110 °C)
-05- HNO

3

(0.00 ... 9.99 % by wt)

(-17 ... 50 °C)

Press enter to proceed

-01-SOL

(-01-SOL

-02-SOL

-03-SOL

-04-SOL

-05-SOL)

Concentration measurement

For the solutions listed above, the transmitter can determine the sub-

stance concentration from the measured conductivity and temperature

values in % by wt. The measurement error is made up of the sum of

measurements errors during conductivity and temperature measurement

and the accuracy of the concentration curves stored in the transmitter,

see page 106.

We recommend to calibrate the transmitter together with the sensor.

For exact temperature measurement, you should perform a temperature

probe adjustment. For measuring processes with rapid temperature

changes, a separate temperature probe with fast response should be

used.

When measuring processes such as dilution or intensification of CIP solu-

tions (Clean-In-Place), it is helpful to switch between the parameter sets

for measuring the process medium and for measuring the CIP solution.