Measurement, Temp probe adjustment – Hach-Lange SI794 C User Manual User Manual
Page 80

80
Display
Remark
Measurement
In the measuring mode the main display shows the
configured process variable (conductivity, resistivity,
or SAL) and the lower display the temperature. During
calibration you can return to measuring mode by
pressing the cal key, during configuration by pressing
conf (waiting time for measured-value stabilization
approx. 20 sec).
Display
Action
Remark
Enter measured temperature
value.
Select position using4key,
edit number using 5key.
Press enter to proceed.
Press enter to exit adjustment.
HOLD will be deactivated after
20 sec.
Temp probe adjustment
Select calibration
(Press cal,
enter passcode: 1015)
Select position using4key,
edit number using 5key.
Press enter to proceed.
Measure the temperature of
the process medium using an
external thermometer
Wrong settings
change the mea-
surement proper-
ties! If an invalid
code is entered, the
transmitter returns
to measuring mode.
Default:
Value of secondary
display.
The transmitter is in
the Hold mode.