Display

Remark

Measurement

In the measuring mode the main display shows the

configured process variable (conductivity, resistivity,

or SAL) and the lower display the temperature. During

calibration you can return to measuring mode by

pressing the cal key, during configuration by pressing

conf (waiting time for measured-value stabilization

approx. 20 sec).

Display

Action

Remark

Enter measured temperature

value.

Select position using4key,

edit number using 5key.

Press enter to proceed.

Press enter to exit adjustment.

HOLD will be deactivated after

20 sec.

Temp probe adjustment

Select calibration

(Press cal,

enter passcode: 1015)

Select position using4key,

edit number using 5key.

Press enter to proceed.

Measure the temperature of

the process medium using an

external thermometer

Wrong settings

change the mea-

surement proper-

ties! If an invalid

code is entered, the

transmitter returns

to measuring mode.

Default:

Value of secondary

display.

The transmitter is in

the Hold mode.