Code display action choices o2, Process temp: current start and current end – Hach-Lange SI794 C User Manual User Manual

Page 47

Advertising
background image

47

°C

(°C / °F)

Code Display

Action

Choices

o2 .

Specify temperature unit

Select using 4 key.

Press enter to proceed.

4 - 20 mA

(4 - 20 mA/

0 - 20 mA)

Set output current range

Select using 4 key.

Press enter to proceed.

Current start: Enter lower end of

scale.

Select using 4 key, edit number

using5 key, press enter to proceed.

000 .0 °C

(xxx.x °C)

Current end: Enter upper end of

scale.

Select using 4 key, edit number

using5 key, press enter to proceed.

100 .0 °C

(xxx.x °C)

Output current

Process temperature

20

4

100

0

Process temp: Current start and current end

Process temperature

20

4

70

50

[mA]

Example 1: Range 0 ... 100 °C

Example 2: Range 50 ... 70 °C.

Advantage: Higher resolution in

range of interest

[°C]

[°C]

[mA]

100 PT

1000 PT

8.55 NTC

30 k NTC

Select temperature probe

Select using 4 key.

Press enter to proceed.

Output current

*) These parameters are only edited in parameter set 1.

They are valid for both parameter sets.