Selecting parameter set 1/2 – Hach-Lange SI794 C User Manual User Manual
Page 66

66
Selecting parameter set 1/2
Manually or via a signal at the Control input
Display
Action
Choices
Select:
• Parameter set 1 (MAN)
• Parameter set 2 (MAN)
• Automatic switchover via
control input (Ctr-EXT)
Select using 4, press enter to proceed.
To select a parameter set, press conf,
enter code: 7654
Select position using 4, edit number
using 5, press enter to proceed.
Wrong settings change the measurement
properties! If an invalid code is entered, the
transmitter returns to measuring mode.
With -1- or -2- selected:
Since the complete device configura-
tion is changed in one step, there is a
security prompt (No/Yes).
Note:
When pressing enter directly,
the selection is not saved.
Activation of parameter set 2 is
indicated by “88” in the upper left
corner of the display.
After correct input a
welcome text (CONF)
is displayed for approx.
3 sec
-1-
(-1- MAN
-2-MAN
Ctr-EXT)
For Ctr-EXT,
see next
page