Diagnostics functions – Hach-Lange SI794 C User Manual User Manual

Page 81

Advertising
background image

81

Display of output currents

Press enter while in measuring mode.

The current at output 1 is shown in the main display,

the current at output 2 in the secondary display.

After 5 sec the transmitter returns to measuring

mode.

Diagnostics functions

Display

Remark

Display of calibration data (Cal Info)

Press cal while in measuring mode and confirm code

0000. The current cell constant is shown in the main

display. After 20 sec the transmitter returns to mea-

suring mode (immediate return at pressing enter).

Sensor monitor

For validation of sensor and complete signal process-

ing. Press conf while in measuring mode and enter

code: 2222. The measured resistance is shown in

the main display, the measuring temperature in the

lower display.

Press enter to return to measurement.

Display of last error message

(Error info)

Press conf while in measuring mode and confirm

code 0000. The last error message is displayed

for approx. 20 sec. After that the message will

be deleted (immediate return to measurement at

pressing enter).

Advertising