Information on calibration

Calibration adapts the transmitter to the conductivity sensor.

Calibration can be performed by:

• Input of cell constant (e.g. for ultrapure-water sensors)

• Determining the cell constant using a known calibration

solution

• Sampling (product calibration)

• Temperature probe adjustment

Note:

• All calibration procedures must be performed by trained

personnel.

• Incorrectly set parameters may go unnoticed, but change

the measuring properties.

Exit

Press cal to exit. The measured

value is shown in the main

display alternately with “Hold”,

“enter” blinks, Sensoface is

active. Press enter key to end

the Hold mode. The measured

value is dispålayed. The output

current remains frozen for

another 20 sec (HOLD icon on,

“hourglass” blinks).