Hach-Lange SI794 C User Manual User Manual
Page 79
Advertising
79
Display
Action
Remark
Measuring mode:
From the blinking CAL mode
indicator you see that product
calibration has not been ter-
minated.
Product calibration step 2:
When the sample value has
been determined, open the
product calibration once more
(cal, code 1105).
Display (approx.
3 sec)
Enter lab value. The new cell
constant is calculated.
Press enter to confirm.
Until the sample
value is deter-
mined and can
be entered, the
transmitter is in
measuring mode.
The new cell constant is
displayed.
Press enter to confirm.
The measured value is shown
in the main display alternately
with “Hold”. “enter” blinks.
Press enter to exit.
After end of
calibration, the
outputs remain
in Hold mode for
approx. 20 sec.
New calibration:
Press cal.