Hach-Lange SI794 C User Manual User Manual
Page 39
39
Code
LIN
(LIN / LOG)
Display
Action
Choices
o1 .
Select output characteristic
Select using 4 arrow key.
Press enter to proceed.
(Step omitted for % (Conc) and
SAL)
000 .0 mS
(xxx.x mS)
100 .0 mS
(xxx.x mS)
Conductivity
[mS/cm]
20
4
200
0
Assignment of measured values:
Current start and current end
[mS/cm]
20
4
200
100
[mA]
Example 1: Range 0...200 mS/cm
Example 2: Range 100...200 mS/cm
Advantage: Higher resolution in range
of interest
[mA]
Output current
Conductivity
4-20 mA
(0 - 20 mA/
4 - 20 mA)
Set output current range
Select using 4 arrow key.
Press enter to proceed
With LIN selected:
• Enter current start
(lower end of scale).
Select position using 4 key,
edit number using 5 key.
Press enter to proceed.
• Enter current end
(upper end of scale).
Press enter to proceed
Output current