Code display action choices o1, Output signal during hold – Hach-Lange SI794 C User Manual User Manual

Page 45

Advertising
background image

45

OFF

(OFF / ON)

Code Display

Action

Choices

o1 .

22 mA signal for error message

Select using 4 key.

Press enter to proceed

Output signal during HOLD

LAST:

During HOLD the last mea-

sured value is maintained at the

output.

FIX:

During HOLD a value (to be

entered) is maintained at the

output.

Select using 4 key.

Press enter to proceed.

LAST

(LAST / FIX)

Only with FIX selected:

Enter current which is to flow at

the output during HOLD

Select position using 4 key and

edit number using5 key.

Press enter to proceed.

021 .0 mA

(00.0 ... 21.0

mA)

Output signal during HOLD:

Output current

[mA]

Output signal for HOLD

FIX setting = 21.0 mA

HOLD active

21

4

Output signal for HOLD

LAST setting

HOLD active

Advertising