Code display action choices o2, Output signal during hold – Hach-Lange SI794 C User Manual User Manual
Page 51
Advertising
51
OFF
(OFF / ON)
Code Display
Action
Choices
o2 .
22 mA signal for error message
Select using 4 key.
Press enter to proceed.
Output signal during HOLD
LAST:
During HOLD the last mea-
sured value is maintained at the
output.
FIX:
During HOLD a value (to be
entered) is maintained at the
output.
Select using 4 key.
Press enter to proceed.
LAST
(LAST / FIX)
Only with FIX selected:
Enter current which is to flow at
the output during HOLD
Select position using 4 key and
edit number using5 key.
Press enter to proceed.
21 .0 mA
(00.0 ...
21.0 mA)
Output signal during HOLD:
Output current
[mA]
Output signal for HOLD
Setting FIX = 21.0 mA
HOLD active
21
4
Output signal for HOLD
LAST setting
HOLD active