Code Display
Action
Choices
o2 .
Time interval of output filter
Default setting: 0 sec (inactive).
To specify a time interval:
Select position using 4,
edit number using 5.
Press enter to proceed.
Time averaging filter (attenuation)
To smoothen the current output 2, a low-pass filter with adjustable
filter time can be switched on. When there is a jump at the input
(100 %), the output level is at 63 % after the time interval has been
reached.
The time interval can be set from 0 to 120 sec.
If the time constant is set to 0 sec (default), the current output
follows the input.
Note:
The filter only acts on the current output, not on the display!
Time inerval 0 to 120 sec