Hach-Lange SI794 C User Manual User Manual

Page 49

Advertising
background image

49

0 sec

(0...120 sec)

Code Display

Action

Choices

o2 .

Time interval of output filter

Default setting: 0 sec (inactive).

To specify a time interval:

Select position using 4,

edit number using 5.

Press enter to proceed.

Time averaging filter (attenuation)

To smoothen the current output 2, a low-pass filter with adjustable

filter time can be switched on. When there is a jump at the input

(100 %), the output level is at 63 % after the time interval has been

reached.

The time interval can be set from 0 to 120 sec.

If the time constant is set to 0 sec (default), the current output

follows the input.

Note:

The filter only acts on the current output, not on the display!

Time inerval 0 to 120 sec