Error messages (error codes), Error display problem possible causes – Hach-Lange SI794 C User Manual User Manual

Page 88

Advertising
background image

88

A

lar

m

con

tac

t

Red

LED

O

ut

1

(22

mA

)

O

ut

2

(22

mA

)

Error messages (error codes)

Sensor

• Wrong cell constant

• Measurement range violation

• SAL > 45 ‰

• Sensor connection or cable defective

Error

Display Problem

Possible causes

ERR 01

Measured

value

blinks

x

x

x

Unsuitable sensor

Conductance range > 3500 mS

ERR 02

Measured

value

blinks

x

x

x

System error

Configuration or calibration data

defective; completely reconfigure

and recalibrate the transmitter.

Memory error in device program

ERR 98

“Conf”

blinks

x

x

x

x

Factory settings

EEPROM or RAM defective

This error message only occurs

in the case of a total defect. The

transmitter must be repaired and

recalibrated at the factory.

ERR 99

“FAIL”

blinks

x

x

x

x

Temperature probe

Open or short circuit

Temperature range exceeded

ERR 03

x

x

x

x