Sensocheck
sensor monitoring
Polarization detection and monitoring of
cable capacitance
Sensoface
Provides information on the sensor condition
(Sensocheck)
Sensor monitor
Direct display of measured values from sensor for
validation (resistance/temperature)
USP function
Water monitoring in the pharmaceutical industry
(USP) with additional user-defined limit value (%),
output via relay contact
Temperature input
*
Pt100 / Pt1000/ NTC 30 kΩ /
NTC 8.55 kΩ (Betatherm)
2-wire connection, adjustable
Measuring range
Pt100 / Pt1000
-20 ... +200 °C
(-4 ... 392 °F)
NTC 30 kΩ
-20 ... +150 °C
(-4 ... 302 °F)
NTC 8.55 kΩ
-10 ... +130 °C
(+14 ... 266 °F)
Resolution
0.1 °C / 1 °F
Measurement error
1,2,3)
0.5 K
(< 1 K for Pt100; <1 K for NTC >100 °C)
Temperature compensation
*
(Reference temp 25 °C)
(OFF) None
(Lin) Linear characteristic 00.00 ... 19.99 %/K
(NLF) Natural waters to EN 27888
(nACL) Ultrapure water with NaCl traces (0...120°C)
(HCL) Ultrapure water with HCl traces (0...120 °C)
(nH3) Ultrapure water with NH
3
traces (0...120 °C)