Hach-Lange SI794 C User Manual User Manual

Page 99

Advertising
background image

99

Sensocheck

sensor monitoring

Polarization detection and monitoring of

cable capacitance

Sensoface

Provides information on the sensor condition

(Sensocheck)

Sensor monitor

Direct display of measured values from sensor for

validation (resistance/temperature)

USP function

Water monitoring in the pharmaceutical industry

(USP) with additional user-defined limit value (%),

output via relay contact

Temperature input

*

Pt100 / Pt1000/ NTC 30 kΩ /

NTC 8.55 kΩ (Betatherm)

2-wire connection, adjustable

Measuring range

Pt100 / Pt1000

-20 ... +200 °C

(-4 ... 392 °F)

NTC 30 kΩ

-20 ... +150 °C

(-4 ... 302 °F)

NTC 8.55 kΩ

-10 ... +130 °C

(+14 ... 266 °F)

Resolution

0.1 °C / 1 °F

Measurement error

1,2,3)

0.5 K

(< 1 K for Pt100; <1 K for NTC >100 °C)

Temperature compensation

*

(Reference temp 25 °C)

(OFF) None

(Lin) Linear characteristic 00.00 ... 19.99 %/K

(NLF) Natural waters to EN 27888

(nACL) Ultrapure water with NaCl traces (0...120°C)

(HCL) Ultrapure water with HCl traces (0...120 °C)

(nH3) Ultrapure water with NH

3

traces (0...120 °C)