Product calibration – Hach-Lange SI794 C User Manual User Manual

Page 78

Advertising
background image

78

For product calibration the measured variable is used as configured:

Conductivity (µS/cm, mS/cm, S/m), resistivity (MΩ·cm). During product

calibration the sensor remains in the process. The measurement is only

interrupted briefly.

Calibration is without TC correction.
Procedure: During sampling the currently measured value is stored in

the transmitter. The transmitter immediately returns to measuring mode.

The calibration mode indicator blinks and reminds you that calibration

has not been terminated.

The sample is measured in the lab or directly on the site using a portable

meter. The sample value is then entered in the transmitter. The new cell

constant is calculated from these two values.

If the sample is invalid, you can take over the value stored during

sampling. In that case the old calibration values are saved. Afterwards,

you can start a new product calibration.

Product calibration

Calibration by sampling

Display

Action

Remark

Product calibration step 1:

Press cal, enter code: 1105.

(Press 4 key to select position,

enter number using 5 key.

Press enter to confirm.)

Take sample and save value.

Press enter to proceed.

If an invalid code is

entered, the trans-

mitter returns to

measuring mode.

The sample is mea-

sured in the lab or

directly on the site.

Display (approx.

3 sec)