Hach-Lange SI794 C User Manual User Manual
Page 74
74
Display
Action
Remark
The lower display
shows the mea-
sured conductivity
value.
(When there has
not been an entry
for 6 sec, the lower
display alternately
shows the conduc-
tivity and tempera-
ture value.)
Calibration by entry of cell constant
Input of cell constant with simultaneous display of uncorrected
conductivity value and temperature
Press cal, enter code: 1100
Select position using 4,
edit number using5,
press enter to proceed.
Enter cell constant of
connected sensor:
Select using 4 key,
edit number using 5 key.
A change in the cell constant
also changes the conductivity
value.
Press enter to confirm cell
constant.
The transmitter is in
the Hold mode.
If an invalid code is
entered, the trans-
mitter returns to
measuring mode.
Ready for calibration
Display (3 sec)