Hach-Lange SI794 C User Manual User Manual

Page 77

Advertising
background image

77

Display

Action

Remark

The determined cell constant is

displayed.

Confirm with enter.

Note:

Be sure to use known calibration solutions with the respective

temperature-corrected conductivity values (see ”Calibration

Solutions” Pg 104).

During the calibration procedure the temperature must be

kept constant.

The transmitter now displays

the conductivity and

temperature.

Clean sensor and re-place it in

the process.

The measured value is shown

in the main display alternately

with “Hold”. “enter” blinks.

End calibration using

enter.

After end of calibra-

tion, the outputs

remain in Hold

mode for approx.

20 sec.