Display
Action
Remark
The determined cell constant is
displayed.
Confirm with enter.
Note:
•
Be sure to use known calibration solutions with the respective
temperature-corrected conductivity values (see ”Calibration
Solutions” Pg 104).
•
During the calibration procedure the temperature must be
kept constant.
The transmitter now displays
the conductivity and
temperature.
Clean sensor and re-place it in
the process.
The measured value is shown
in the main display alternately
with “Hold”. “enter” blinks.
End calibration using
enter.
After end of calibra-
tion, the outputs
remain in Hold
mode for approx.
20 sec.