OFF
(OFF
LIN
nLF
nACL
HCL
nH3)
Code Display
Action
Choices
tc .
Only with linear temperature
compensation (LIN) selected:
Enter temperature coefficient.
Select position using 4 key and
edit number using 5 key.
Press enter to proceed.
02 .00%/K
(XX.XX %/K)
Select temp compensation
(not for USP)
OFF: Temperature compensation
switched off
Select using 4key.
Press enter to proceed.
LIN:
Linear temperature compensa-
tion with entry of temperature
coefficient and reference
temperature
nLF:
Temperature compensation for
natural waters to EN 27888
NaCl (nACL):
Temperature compensation for
ultrapure water with NaCl traces
HCl (HCL):
Temperature compensation for
ultrapure water with HCl traces
NH3 (nH3):
Temperature compensation for
ultrapure water with NH
3
traces