Signaling parameter set 1/2

17

18

Clean

e.g. spray cleaning

Power supply

Controlling a rinsing probe

The “Clean” contact can be used to connect a simple rinsing probe.

Rinse duration and rinsing interval are defined during configuration.

Contact response can be set as N/O or N/C.

Parameter set 1 selected

Parameter set 2 selected

17

18

Clean

PSet2

Process control system

Power supply

si794

Power supply:

AC< 250 V / < 3 A / < 750 VA

DC< 30 V / < 3 A / < 90 W

Depending on the selected parameter set, the relay

is active or inactive. The signal can be used for super-

ordinated process control systems.

Parameter set 2 is indicated by “88” in the upper left

corner of the display.