Beamage-3.0 User Manual Revision 6.0

28

4.5.3. Trigger

Figure 4-30 Trigger Button

The

“Trigger” button enables the camera to capture images only when an electric signal is sent to the

Beamage-3.0

via the SMA connector. This can synchronize the system’s capture rate with a pulsed laser

source. A SMA to BNC adaptor is provided (202273). The input trigger signal can be from 1.1 volts to 24
volts. The rise edge response time is 300 ns. The pulse width of the trigger signal must be between 300
ns and 230 ms.

Figure 4-31 SMA Connector for Trigger Input

4.5.4. Show/Hide Options

Figure 4-32 Show/Hide Options Button

The

Show/Hide Options” button will show or hide the “Divergence”, the “Relative Position”, “Camera Lens

Calibration

” and “Fixed Crosshair” panels beside the “Data Acquisition” tab. By clicking on “Show All’ or

“Hide All”, one can show or hide both tabs at the same time. The “Start LabVIEW Pipeline” button will
open the communication channel between the PC-Beamage-3.0 and the LabVIEW driver. Please refer to
section 7 for more information about this function.

Figure 4-33 Show/Hide Options

SMA Connector