5.2.4. Active Area

The

“Active Area” function allows the user to select a region of interest (ROI) on the sensor. This will

increase the frame rate, as fewer pixels need to be transferred from the Beamage-3.0. This can only be
done on small beam sizes, since a cropped beam would invalidate the beam width measurements.
Furthermore, to have an accurate measurement, the active area must be at least 2 times the beam size.

The user can select the desired area from a preset selection or enter a custom size. By default, the area
will be placed at the sensor’s upper left corner pixel (0, 0). This position can be changed by entering the
active area’s upper left position. Checking the “Center” check box will center the active area to the
sensor’s center.

Tip

When working with small beams, optimize the speed of the data

transfer and still maintain accurate results, by using an Active Area that is

twice the size of your beam.

5.2.5. Pixel Addressing

The

“Pixel Addressing” mode allows the user to downsample the captured image. The “Average 2x2” will

take a 2x2 pixel cluster and return its average as one larger pixel.

The “Average 2x2” function is only

available in the 12 bit ADC mode. The

“Decimate 2x2” will only return 1 out of the 4 pixels. Because the

pixel area is doubled with this mode, it can be used with large beams, where the spatial resolution is not
crucial. It will increase the frame rate because fewer pixels are transferred from the Beamage-3.0.

Full Resolution

Average 2x2

Decimate 2x2

Figure 5-7 Pixel Addressing Mode

Tip

When working with large beams, optimize the speed of the data transfer

by reducing the spatial resolution using the Pixel Addressing function.


Advertising