Amera, Igure, Alibration – Gentec-EO Beamage-3.0 User Manual
Page 48: Ection, Oving, Irection
Beamage-3.0 User Manual Revision 6.0
47
5.6.
CAMERA LENS
Prior to profiling a beam with a camera lens, one must adjust the Pixel Magnification Factor of the lens
(see section 5.2.8).
The
“Camera lens calibration” section allows the user to calibrate the PC-Beamage software when a
camera lens is used with the Beamage-3.0.
This panel is accessible by clicking “Calibrate” in the Pixel
Multiplication factor section in the Setup panel or in the Show/Hide Options in the Ribbon.
Figure 5-15 Camera Lens Calibration Section
1. Set up the laser and the camera lens with the Beamage-3.0.
2. Click on Set now to set the centroid to the current position
3. Then, move the laser source (or the Beamage-3.0 camera) by a known distance along the X axis,
parallel to the diffuser.
Figure 5-16 - Camera Lens Calibration Moving Direction