Implanted pacemakers, Lead monitoring – ZOLL R Series Monitor Defibrillator Rev K Operators Guide User Manual

Page 114

Advertising
background image

C

HAPTER

10

ECG M

ONITORING

10–6

www.zoll.com

9650-0904-01 Rev. K

WARNING!

Implanted pacemakers may cause the heart rate meter to count pacemaker pulses
during incidents of cardiac arrest or other arrhythmias. Pacemaker patients should be
carefully observed. Check the patient's pulse; do not rely solely on heart rate meters.
Dedicated pacemaker detection circuitry may not detect all implanted pacemaker
spikes; patient history and physical exam are important in determining the presence of
an implanted pacemaker.

Implanted Pacemakers

The R Series device can be configured to detect pacemaker signals from a patient with an
implanted pacemaker and indicate their presence on the display. When pacer pulses are
detected, the device displays a 5mm, vertical, solid line on the ECG trace.

To disable detection of pacemaker spikes:

1. Press the Param softkey.

2.

Press ECG.

3.

Press Disable Pacer Detect.

To re-enable detection of pacemaker spikes:

1. Press the Param softkey.

2.

Press ECG.

3.

Press Enable Pacer Detect.

5-Lead Monitoring

You can perform 5-lead ECG monitoring with the appropriate ECG patient cable. The 5-lead
cable allows you to monitor the following ECG leads:

I, II, III

aVR, aVL, aVF

V1

Param

Report

Data

Test

Log

72

ECG

PADS

x1

SpO2%

0:00

IDLE

Depth
Rate

99

NIBP

mmHg

---
---

---

CO2

mmHg

---
---

RR

Ɔ

Options

Code

Marker

ECG

MONITOR

Alarms

CPR

PPI

Pace Marker

Advertising