Transferring Device Check and Activity Log Files to a
Compact Flash Card

Make sure that a compact flash card is installed in the slot on the left side of the device.

To copy the Device Check or Activity Log to a CF card:

1. Turn the Mode Selector to ON.

2.

Press the Manual Mode softkey.

Note:

If the CONFIRM MANUAL MODE message displays, press the Confirm softkey to
enter the Manual mode of operation.

3.

Press the Report Data softkey.

4.

Press the Transfer Mode softkey.

5.

Press the More softkey.

6.

Press the Defib History to Card softkey.

The message TRANSFERRING DATA is displayed. The activity log and test file is transferred
to the installed CF data card.

Note:

If the messages CF TRANSFER FAILED and then TRANSFERRING DATA display,
make sure you have a CF card installed and not a Wi-Fi card. Press the Return softkey
and then the Exit Transfer softkey to return to defibrillation operation mode.

Note:

Do not remove the CF card while files are transferring. Corruption may result on the
data card.

When all files are transferred, the message DATA TRANSFERRED is displayed. You can now
remove the CF card from the R Series unit.

To exit Data Transfer Mode, press the Return softkey and then the Exit Transfer softkey.

Transferring Files Through the USB Port (R Series Plus Only)

Before you begin, connect a USB cable from the Windows external device with USB host
capability (for example, a Window PC), to the R Series defibrillator USB device port.

Figure 12-3.

USB Ports

USB Device Port

USB Host Port
(Future Use)