background image

Defibrillator

9650-0904-01 Rev. K

ZOLL R Series Operator’s Guide

14–5

Defibrillator

4. Patient on “Standby” pacing gets

paced intermittently.

Ensure proper ECG electrode or OneStep Pacing/
Complete electrode connection and placement. If ECG
lead wire comes off, pacer will automatically pace
asynchronously.

Check ECG cable for damage.

Patient R-to-R interval varying. Pace rate close to
patient’s heart rate.

Verify rate is set appropriately.

5. Heart rate display reads 0 with

proper pacing capture displayed
on ECG trace.

Check patient’s pulse.

Select different ECG Lead configuration.

6. Bedside/Central Station/Telemetry

ECG display becomes erratic
when pacing.

None, the patient monitor ECG inputs are overloaded by
pacer signals. ECG can only be monitored by the R Series
or pacing device while pacing.

Symptom

Recommended Action

1. Excessive artifact detected when

using paddles as ECG source.

Ensure “PADDLES” is selected.

Firmly press paddles against patient skin.

Use gel on paddles.

Clean paddle surface.

Check and clean between adult and pediatric shoe.

Check cable for damage.

Use ECG electrodes.

2. Defibrillator will not charge (energy

level does not increment on
display).

Ensure that SHOCK button(s) on paddles or front panel
are not stuck on.

Replace the battery pack.

3. Charge time to 200 J exceeds

10 seconds.

Typical in a low battery condition (up to 20 seconds)

Plug device into AC power.

Install fully charged battery pack.

Symptom

Recommended Action