ZOLL R Series Operator’s Guide

9–1

Chapter 9

Noninvasive Temporary Pacing

(Optional)

Note:

This is an optional feature and is only available in Manual mode.

WARNING!

To avoid risk of electrical shock, do not touch the gelled area of the hands-free therapy
electrodes while pacing.

Therapy electrodes should be replaced periodically. Consult electrode directions for
specific recommendations.

Prolonged pacing (in excess of 30 minutes), particularly in neonates or adults with
severely restricted blood flow, may cause burns. Periodic inspection of the underlying
skin is recommended.

If the unit was NOT turned off and less than 10 minutes have elapsed since the pacing
mode was last used, reactivating the pacer mode causes pacing to resume immediately
at the previously selected mA and ppm settings.

When ZOLL hands-free therapy electrodes are used, the patient connection is
considered to be defibrillation-protected Type BF.

ECG leads are a defibrillation-protected Type CF patient connection.