1 select on – ZOLL R Series Monitor Defibrillator Rev K Operators Guide User Manual

Page 86

Advertising
background image

C

HAPTER

6

S

YNCHRONIZED

C

ARDIOVERSION

6–6

www.zoll.com

9650-0904-01 Rev. K

Determine the Patient’s Condition and Provide Care Following Local
Medical Protocols

Prepare Patient

Prepare the patient as described in “Prepare Patient” on page 6-2.

Follow the instructions provided with the external monitoring device to prepare the patient for
ECG monitoring and synchronization with a separate defibrillator.

1 Select ON

The unit beeps 4 times to indicate that it has passed the power on self test and the green AED
label and the ON mode selector light up on the front panel.

Press the Manual Mode softkey on the front panel of the unit to enter the Manual mode of
operation.

The message CONFIRM MANUAL MODE and the Confirm softkey will be displayed. Press
the Confirm softkey to enter the Manual mode of operation. If the Confirm softkey is not
pressed within 10 seconds, the unit will revert back to AED operation.

Note:

If the unit has been configured not to confirm the transition from AED mode to
Manual mode, the unit will immediately enter Manual mode.

The control buttons for Manual mode light up on the front panel. Select the desired energy
using the up and down arrow keys on the front panel (or sternum paddle if using paddles).

Press Sync On/Off softkey, then press the Remote Sync Softkey

The selected energy level is displayed on the monitor.

The words “REMOTE SYNC” are displayed in place of the ECG trace, and a REMOTE SYNC
XXXJ SEL.
message appears on the display.

The ECG heartbeat indicator will flash with each synchronization pulse received from the
remote monitoring device.

Unless otherwise configured, the unit automatically exits Sync mode after each shock, and if
the Mode Selector is moved to OFF.

Press the Sync On/Off, Remote Sync softkey sequence again to reactivate Remote Sync mode.
Changing the selected energy levels does not cause the unit to leave Remote Sync mode.

OFF

ON

AED