Begin cpr following local medical protocols, 1 select on – ZOLL R Series Monitor Defibrillator Rev K Operators Guide User Manual

Page 62

Advertising
background image

C

HAPTER

4

M

ANUAL

D

EFIBRILLATION

4–2

www.zoll.com

9650-0904-01 Rev. K

Begin CPR Following Local Medical Protocols

Request additional assistance.

1 Select ON

The unit beeps 4 times to indicate that it has passed the power on self test, and the green AED
label and the ON mode selector light up on the front panel.

Press the Manual Mode softkey on the front panel of the unit to enter the Manual mode of
operation.

The message CONFIRM MANUAL MODE and the Confirm softkey will be displayed. Press
the Confirm softkey to enter the Manual mode of operation. If the Confirm softkey is not
pressed within 10 seconds, the unit will revert back to AED operation.

Note:

If the unit has been configured not to confirm the transition from AED mode to
Manual mode, the unit will immediately enter Manual mode.

The control buttons for Manual mode light up on the front panel and the unit automatically
defaults to 120 joules or the preconfigured first shock energy selection.

Note:

Defibrillator PADDLES are selected as the ECG source when the instrument is
changed to Manual mode with paddles connected to the OneStep cable.

Energy Select

Look at the Display and verify the energy is appropriate. Unless internal handles are connected
to the OneStep cable, the default energy selections for adult patients are:

Shock 1 - 120 joules

Shock 2 - 150 joules

Shock 3 - 200 joules

OFF

ON

AED