Summary Report

9650-0904-01 Rev. K

ZOLL R Series Operator’s Guide

11–5

Heart Rate Alarm Activated Format

The summary report function records 6 seconds of pre-alarm patient ECG. Also recorded are
the ECG lead, ECG size, patient's heart rate, actual event time, and date. The date/time printed
on the top of the strip corresponds to the ECG data occurring 6 seconds before the event. If the
pacer is on during this event, the pacing rate and pacing current are also recorded.

Figure 11-6.

Heart Rate Alarm Activated Format

VF Alarm Activated Format

The summary report function records 18 seconds of patient ECG data associated with each VF
alarm. Also recorded are the ECG lead, ECG size, actual event time, the number of noise
events, and the message CHECK PATIENT. The date/time printed on the top of the strip
corresponds to the ECG data occurring 6 seconds before the event.

Figure 11-7.

VF Alarm Activated Format