Transferring Device Check and Activity Log Files Through Wi-Fi (R Series Plus Only)

9650-0904-01 Rev. K

ZOLL R Series Operator’s Guide

12–9

TRANSFER FAILED
Invalid Wi-Fi configuration
Error Number: N
Contact Network Administrator

(cont.)

N = 8: The Defib Upload Server's Username /
Password are blank within the Full Disclosure and Defib
History server configuration items Data COMM II only)

Action:

If Full Disclosure Transfer Mode is set to Server, make
sure that the Username (Full Disclosure) and Password
(Full Disclosure) entries are valid.

If Defib History Transfer Mode is set to Manual,
Automatic, or Both, make sure that the Username
(Defib History) and Password (Defib History) entries are
valid.

N = 9: Missing required certificate file for Network
Profile 1 with Enterprise Authentication selected
(Data COMM II only)

Action:

If Authentication Protocol is set to PEAP, make sure
that a root certificate has been added (if not using the
default ZOLL root certificate).

If Authentication Protocol set to TLS, make sure that a
root certificate has been added (if not using the default
ZOLL root certificate) and that the correct client
certificate has been configured.

N = 10: Missing required certificate file for Network
Profile 2 with Enterprise Authentication selected
(Data COMM II only)

Action: Same as above.

TRANSFER FAILED
Network Error: General

Error Number: NNNN
Contact Network Administrator

The transfer failed because of a general network Wi-Fi
error. NNNN may indicate one of the following error
codes:

N = 7004
N = 7010
N = 7011
N = 7015
N = 7016
Action: Install another ZOLL Data COMM card.

N = 7500
N = 7501
N = 7502
N = 7503
N = 7504
N = 7505
N = 7506
N = 7507
Action: Install another Data COMM II card.

N = 7017
Action: Make sure that the Defib Upload Server is
running, is operational, and can be reached on the
network.