Operator’s Checklist for R Series Product

Recommended checks and procedures to be performed
daily or weekly, depending on Automatic Test
configuration.

Date

______________________________________________

Location ____________________________________________

Unit Serial Number ___________________________________

1. Condition

Remarks

Unit clean, no spills, clear of objects on top, case intact

2. Hands-free Therapy electrodes

1 set preconnected / 1 spare

3. Paddles

Paddles clean, not pitted

Release from housing easily

4. Inspect cables for cracks, broken wires, connector

A ECG electrode cable, connector

B Defibrillator paddle cables

C OneStep cable, connector

D Other patient cables

5. Batteries

A Fully charged battery in unit

B Fully charged spare battery available

6. Disposable supplies

A Defib gel or gel patches

B Hands-free therapy electrodes in sealed packages — 2 sets

C ECG electrodes

D Recorder paper

E Alcohol wipes

F Razors

7. Operational checks

A Power On Sequence

Turn unit to ON, 4-beep tone heard, AED label and ON Mode Selector light up

Press Manual Mode softkey, press Confirm softkey, front panel lights up

ECG size X 1

“PADDLES” or ‘PADS” as lead selected

B Paddles

Paddles in holder: Set defib energy level to 30 joules, press paddles firmly into the side

wells, and simultaneously press and hold both defib discharge buttons; “30J TEST OK”

message on Recorder.

C Defibrillator

OneStep cable connected to test connector,

or OneStep electrodes.

Set defib energy

level to 30 joules, press SHOCK button; “30J TEST OK” message on Recorder

D Pacer Operation (Optional)

OneStep cable not connected to Test Connector or OneStep electrodes

Turn to PACER, set pacer rate to 150 ppm, press RECORDER button

Pacer pulses occur ever 2 large divisions (10 small divisions)

Press 4:1 button, pulses occur every 8 large divisions

Set PACER OUTPUT to 0 mA, no CHECK PADS message

Set PACER OUTPUT to 16 mA, CHECK PADS message and alarm

Reconnect OneStep cable to test connector,

or OneStep electrodes

. Press Clear Pace

Alarm softkey; CHECK PADS message disappears and Pace alarm stops.

E Recorder

Press RECORDER button; Recorder runs. Press again; Recorder stops.

Inspect Recorder printing

8. Please check the appropriate box after each use of this checklist.

Signature

No action required

______________________

Minor problem(s) corrected

Disposable supplies replaced

Major problem(s) identified — UNIT OUT OF SERVICE