Electromagnetic Immunity Declaration (EID)

Guidance and manufacturer’s declaration — electromagnetic immunity for the R Series.

The R Series is intended for use in the electromagnetic environment specified below. The customer or user of
the R Series should ensure that it is used in such an environment.

Immunity test

IEC 60601 test level

Compliance level

Electromagnetic environment –
guidance

Electrostatic
discharge (ESD)
IEC 61000-4-2

±6 kV contact
±8 kV air

±6kV contact
±8 kV air

Floors should be wood, concrete, or
ceramic tile. If floors are covered with
synthetic material, the relative
humidity should be at least 30%.

Electrical fast
transient/burst
IEC 61000-4-4

±2 kV for power supply
lines
±1 kV for input/output
lines

±2 kV for power
supply lines
Not applicable

Mains power quality should be that of
a typical commercial or hospital
environment.

Surge
IEC 61000-4-5

±1 kV differential mode
±2 kV common mode

±1 kV differential
mode
±2 kV common mode

Mains power quality should be that of
a typical commercial or hospital
environment.

Voltage dips,
short interruptions
and voltage
variations on
power supply
input lines
IEC 61000-4-11

<5% U

T

(>95% dip in U

T

)

for 0.5 cycle
40% U

T

(60% dip in U

T

)

for 5 cycles
70% U

T

(30% dip in U

T

)

for 25 cycles
<5% U

T

(>95% dip in U

T

)

for 5 seconds

<5% U

T

(>95% dip in

U

T

) for 0.5 cycle

40% U

T

(60% dip in

U

T

) for 5 cycles

70% U

T

(30% dip in

U

T

) for 25 cycles

<5% U

T

(>95% dip in

U

T

) for 5 seconds

Mains power quality should be that of
a typical commercial or hospital
environment. If the user of the
R Series requires continued operation
during power mains interruptions, it is
recommended that the R Series be
powered by an uninterruptible power
supply or a battery.

Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8

3 A/m

3 A/m

Power frequency magnetic fields
should be at levels characteristic of
typical location in a typical commercial
or hospital environment.

NOTE: U

T

is the AC mains voltage prior to the application of the test level.