ZOLL R Series Monitor Defibrillator Rev K Operators Guide User Manual

Page 88

Advertising
background image

C

HAPTER

6

S

YNCHRONIZED

C

ARDIOVERSION

6–8

www.zoll.com

9650-0904-01 Rev. K

Once the energy is delivered, the display simultaneously shows XXXJ DELIVERED and DEFIB
XXXJ SEL
. After approximately 5 seconds, the XXXJ DELIVERED message disappears and the
DEFIB XXXJ SEL. message remains to indicate the selected energy level.

If additional countershocks are necessary, readjust the energy level as necessary, press the
Sync On/Off, and then the Remote Sync softkeys and repeat. Note that REMOTE SYNC XXXJ
SEL
must be displayed prior to pressing the CHARGE button.

If the ANALYZE button is pressed while the unit is in Remote Sync mode, the unit displays the
REMOVE SYNC message and disallows ECG rhythm analysis until the unit is taken out of Sync
mode.