background image

C

HAPTER

14

T

ROUBLESHOOTING

14–6

www.zoll.com

9650-0904-01 Rev. K

4. Energy does not discharge when

the SHOCK button(s) is pressed.

60 or 120 seconds (user configurable) have elapsed in
Manual mode since initial charge ready. Energy was
internally discharged.

Device is in Sync mode or Remote Sync mode and no
QRS complex is detected.

Energy internally discharged because energy selection
was changed during charge or after the device was
ready.

Unit was not completely charged when SHOCK
button(s) were pressed. Wait for DEFIB XXXJ READY
message and ready tone.

Press and hold SHOCK button(s) until energy is
delivered to the patient.

Pads or paddles not making good contact with patient.

5. Unable to SHOCK when in Sync

mode

Ensure SYNC XXXJ SEL is displayed on monitor.

Make sure ECG signals are displayed.

Check for Sync markers (arrow above R wave). If not
present, change lead selection, or electrode placement.

Press and hold SHOCK button(s) until energy is
delivered to the patient.

Alter ECG electrode placement.

6. Unable to SHOCK when in

Remote Sync mode

Ensure REMOTE SYNC is displayed in place of the
ECG waveform and that REMOTE SYNC XXXJ SEL. is
displayed.

Ensure that a remote device conforming with the Sync
In/Marker Out specifications in Appendix A is properly
connected.

Ensure that Sync markers appear with each R-wave on
the remote device’s display. If Sync markers are not
present on the remote device’s display or do not appear
to be nearly simultaneous with each R-wave, do not
proceed with synchronized cardioversion.

Press and hold SHOCK button(s) until energy is
delivered to the patient.

7. No apparent energy delivery to

patient

Under certain circumstances, some patients will not
show a physical reaction when energy is delivered.

Perform defibrillator self-test.

Check for CHECK PADS and POOR PAD CONTACT
messages alternating on the monitor.

If hands-free therapy electrodes are used, ensure
proper placement and contact.

Review the discharge stripchart for Joules/current
delivered.

Symptom

Recommended Action