Battery pack specifications – ZOLL R Series Monitor Defibrillator Rev K Operators Guide User Manual

Page 168

Advertising
background image

A

PPENDIX

A

S

PECIFICATIONS

A–6

www.zoll.com

9650-0904-01 Rev. K

Battery Pack Specifications

Printing method

High-resolution, thermal array print head

Printout modes

Manual or automatic; user-configurable

Data card

Type

Compact flash card

Sync in / Marker out / ECG out

Sync In

0-5 V (TTL Level) pulse, active high, 5 to
15 msec in duration, no closer than 200
ms apart. Energy transfer begins within
25 ms of the leading edge of the external
sync pulse.

0-5 V (TTL Level) pulse, active high, 10
ms in duration, the leading edge of the
pulse occurs within 35 ms of the peak of
the R-wave)

1.0 V/cm of deflection on stripchart
recorder

<25 ms delay from patient ECG input

Marker Out

ECG Out

Wi-Fi Card

ZOLL R Series
Data COMM

Model: 802.11 abgn Wireless LAN Compact Flash Card

GND

GND

ECG
OUT

MARKER

SYNC
IN

R Series Connector

Pinout

OUT

SWITCHCRAFT CONXALL
17982-7SG-300
Mating Connector:
SWITCHCRAFT CONXALL
16982-7PG-522

Type

Rechargeable lithium ion

Weight

1.7 lb (0.77 kg)

Nominal voltage

10.6 V

Recharge time

4 hours or less within R Series.

Operating time

For a new, fully charged battery at 20°C:

100 defibrillator discharges at maximum energy (200 joules), or

4 hours of continuous ECG monitoring, or

3.5 hours of continuous ECG monitoring and pacing at 60 mA, 80
pulses per minute

Low battery indicator

The message LOW BATTERY is displayed on the screen when there
is approximately 15 minutes of ECG monitoring time left on the
battery. Two-beep low battery tone sounds once a minute until just
before shutdown when the unit beeps twice every 2 seconds.

The time from display of the message LOW BATTERY or REPLACE
BATTERY
until the defibrillator shuts down varies depending on the
battery age and condition.

Battery Shelf Life

3 months before retest and recharge