PECIFICATIONS

A–14

www.zoll.com

9650-0904-01 Rev. K

* For all energy levels, accuracy is equal to either ±15% or 3 joules, whichever is greater.

The R Series Rectilinear Biphasic waveform employs the same first and second phase timing,
similar first and second phase currents/voltages, and essentially the same mechanisms for
controlling defibrillation waveshape as the ZOLL M Series

®

defibrillator. The M Series and

R Series defibrillation waveforms are considered substantially equivalent.

Figures A-1 through A-21 show the Rectilinear Biphasic waveforms that are produced when
the R Series defibrillator is discharged into loads of 25, 50, 75, 100, 125, 150, and 175 ohms at
each energy setting.

The vertical axis shows the current in amperes (A); the horizontal axis shows the duration of
time in milliseconds (ms).

Figure A-1.

Rectilinear Biphasic Waveform at 200 Joules

150 J

107 J

164 J

180 J

183 J

188 J

184 J

179 J

±15%

200 J

142 J

230 J

249 J

253 J

269 J

261 J

260 J

±15%

Table A-2.

Delivered Energy at Every Defibrillator Setting into a Range of Loads

Selected

Energy

Load

25

50

75

100

125

150

175

Accuracy*

-40

-30

-20

-10

0

10

20

30

40

0

2

4

6

8

10

12

ms

A

25 Ohm

50 Ohm

75 Ohm

100 Ohm

125 Ohm

150 Ohm

175 Ohm

200 Ohm