ZOLL R Series Monitor Defibrillator Rev K Operators Guide User Manual

Page 124

Advertising
background image

C

HAPTER

11

E

VENT

R

ECORDS

AND

R

EPORTS

11–4

www.zoll.com

9650-0904-01 Rev. K

Pacer Mode Selected Format

The summary report function records 6 seconds of pre-pace patient ECG data. Also recorded
are the ECG lead, ECG size, patient's heart rate, pace rate, pace current, time and date. The
date/time printed on the top of the strip corresponds to the ECG data occurring 6 seconds before
the event.

Figure 11-4.

Pacer Mode Selected Format

After establishing a paced rhythm, turning the recorder on briefly records the paced rhythm for
later reports. If async pace is active, the annotation ASYNC PACE is also recorded and printed.
The date/time printed on the top of the strip corresponds to the ECG data occurring 6 seconds
before the event.

Figure 11-5.

Pacer Mode Selected Format (Asynchronous Pacing)

Advertising