Advisory Defibrillation

When the Mode Selector is turned to ON, the unit is in Manual mode, and hands-free therapy
electrodes are used, the R Series can identify shockable rhythms using its built in ECG analysis
capability. You must read the advisory messages, charge the defibrillator to the preconfigured
or user-selected energy level (if automatic charge is disabled), and deliver treatment to the
patient when required by protocol and patient condition.

The advisory function can be activated only when:

Hands-free therapy electrodes are connected and selected as the ECG source.

Hands-free therapy electrodes are properly connected to the patient.

The Mode Selector is turned to ON and the unit is in Manual mode.

WARNING!

Use only pediatric electrodes to defibrillate patients under 8 years of age in Advisory
mode. Use of adult electrodes with pediatric patients can result in the delivery of
excessive energy doses.

ZOLL hands-free therapy electrodes are a defibrillation-protected Type BF patient
connection.