IEC 60601-1-2 Specifications

9650-0904-01 Rev. K

ZOLL R Series Operator’s Guide

A–7

IEC 60601-1-2 Specifications

This section provides specification tables for the R Series as per IEC 60601-1-2.

Electromagnetic Emissions Declaration

Guidance and manufacturer’s declaration — electromagnetic emissions for the R Series.

The R Series is intended for use in the electromagnetic environment specified below. The customer or user of
the R Series should ensure that it is used in such an environment.

Emissions test

Compliance

Electromagnetic environment – guidance

RF Emissions
CISPR 11

Group 1

The R Series uses RF energy for its internal function only.
Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.

RF Emissions
CISPR 11

Class B

The R Series is suitable for use in all establishments, including
domestic establishments and those directly connected to the
public low-voltage power supply network that supplies buildings
used for domestic purposes.

Harmonic Emissions
IEC 61000-3-2

Class A

Voltage
Fluctuations/
Flicker Emissions
IEC 61000-3-3

Complies