background image

Manual Defibrillator Testing

9650-0904-01 Rev. K

ZOLL R Series Operator’s Guide

13–11

Figure 13-6.

Inserting the Paper and Closing Printer Door

After the paper is loaded, press the RECORDER button to resume printing.

Note:

To ensure you have loaded the paper properly, check to make sure the arrows printed
on the red, grid side of the paper point upward.

Cleaning the Print Head

To clean the recorder print head, perform the following steps:

1. Press the release button, and allow the printer door to open (see Figure 13-5); then remove

any paper.

2.

Locate the printhead along the front floor of the printer compartment, just below the release
button.

3.

Gently wipe the printhead with a cotton swab moistened with isopropyl alcohol, and dry any
residual alcohol with another dry cotton swab.

Figure 13-7.

Cleaning the Print Head

4.

Place the paper back into the unit and close the cover (see Figure 13-6).